ALERT!
This site is not optimized for Internet Explorer 8 (or older).
Please upgrade to a newer version of Internet Explorer or use an alternate browser such as Chrome or Firefox.
Protocol: Temporary Pacemaker Verification and Troubleshooting
Friday, August 28, 2015
Ley J. Protocol: Temporary Pacemaker Verification and Troubleshooting. October 2022. doi:10.25373/ctsnet.21291060.v1
Determining the Stimulation Threshold
- NOTE: a paced rhythm must be present and patient stable
- From 100% pacing, gradually decrease output for A or V, while watching monitor until you lose capture
- Now gradually increase output until 1:1 capture returns – this is the stimulation threshold
- This is the lowest amount of energy that will reliably capture the heart and pace
- Set mA 1.5-3 times above this threshold value for safety
- Repeat for other circuit (A or V)
- Note: the stimulation threshold should be checked DAILY by the MD/NP when pacing is in use; in the event the threshold value is rising, discuss with surgeon to determine if alternative pacing system is needed
Determining the Sensing Threshold
- Note: the patient must have an underlying rhythm and tolerate a brief period without pacing to perform
- Set RATE at least 10 bpm below patient’s intrinsic heart rate (to allow sensing)
- Set OUTPUT at 0.1 mA (to avoid competitive pacing when device is made asynchronous briefly)
- Highlight SENSITIVITY (under Menu 1)
- Decrease SENSITIVITY: Slowly turn dial counter-clockwise until pace indicator flashes regularly (asynchronously)
- Increase SENSITIVITY: Slowly turn dial clockwise until sense indicator flashes again (when sensing resumes, this is the sensing threshold – the smallest complex that is reliably recognized, thus inhibiting the device)
- Set SENSITIVITY < half this threshold value for safety
- Repeat for other circuit (A or V)
- Restore previous rate and output values
Rapid Atrial (Overdrive) Pacing – PERFORMED BY MD/NP/CNS ONLY
- Used to “overdrive pace” and terminate some atrial reentrant tachyarrhythmias, especially atrial flutter
- ALWAYS VERIFY correct wire connections to avoid stimulating the ventricles at a rapid rate!
- Set maximal atrial output to promote capture (20 mA)
- Go to RAP menu (press MENU button three times)
- Dial in RAP rate 10-40 bpm above intrinsic atrial rate (default = 320)
- From RAP menu: press and hold SELECT button
- Rapid pacing will be delivered via atrial circuit
- Assess monitor for atrial “capture” at rapid rate
- Abruptly discontinue pacing by releasing SELECT button
- Examine rhythm and repeat if unsuccessful, consider: increasing rate and/or duration of stimulus, or gradually reducing pacing rate while holding SELECT
Disclaimer
The information and views presented on CTSNet.org represent the views of the authors and contributors of the material and not of CTSNet. Please review our full disclaimer page here.